Perkin
Elmer Hydrogen Thyratrons The Perkin
Elmer thyratrons are high energy switches capable of operation up to 20
kA and 75 kV. A wide range of standard thyratrons are offered, all constructed
of rugged ceramic and metal parts. These tubes are typically used in applications
such as gas laser, radar, and other modulator applications. Five basic
sizes from I to 41/2 inches in diameter define the thyratron types in
production at Perkin Elmer.
Design
Features
The basic thyratron is a three electrode, low pressure gas filled vacuum
tube, with a thermionically emitting cathode (See Figure 8). It is a
triggerable, closing only electronic switch. This basic design is used
in the vast majority of applications. However, additional features have
been added in order to enhance performance in certain situations.
Auxiliary Grid (TETRODE)
Basic thyratrons usually require several hundred nanoseconds to switch
"on" (time measured from a standard fixed level on the rise
of the trigger pulse until a fixed level is reached on the rise of the
current discharge waveform). Addition of an auxiliary grid can substantially
reduce the switch-on delay time (Figure 9). In actual usage, the switch-on
delay time is inversely related to the current being driven in the auxiliary
grid-cathode circuit until a current level is reached where little or
no further reduction of the turn on delay time is realized.
The auxiliary grid may be run in the "DC primed," "pulsed,"
or "combination" modes. The most common mode is the "DC
primed" mode. In this mode, a DC voltage is applied between the
auxiliary grid (+) and the cathode (chassis ground). There is usually
a series resistor to limit the DC current in the auxiliary grid circuit.
When the auxiliary grid is operated in the "pulsed" mode,
a positive (with respect to the cathode) pulse is applied to the auxiliary
grid which draws a current that can be up to an order of magnitude greater
than would be the case for the "DC primed" example. The pulse
is applied to the auxiliary grid about I microsecond prior to pulsing
the control grid. The auxiliary grid pulse may end at any time after
the control grid pulse has been on for at least I microsecond.
Operating the auxiliary grid in the "combination" mode combines
both "DC primed" and "pulsed" modes. In all cases,
gas is ionized between the cathode and the auxiliary grid when the control
grid is pulsed. Therefore, the time required to ionize the gas and establish
electron flow to the auxiliary grid level is already past, leaving only
the time required to spread the plasma from the auxiliary grid level
through the control grid level. The total time from application of the
control grid pulse to the start of main discharge current flow is substantially
reduced.
Liquid Cooling
For several reasons, it is desirable for the tube's electrodes to achieve
at least certain minimum temperatures during operation. However, where
a high anode heating factor and/or high RMS current is required in the
application, excessive heat may develop. In these cases liquid cooling
the thyratron may allow operation at those otherwise unacceptable conditions.
If liquid cooling is not possible or sufficient, the next larger tube
size is recommended. Two methods of liquid cooling the tube are in general
use, immersion and circulation. In some cases, the desired performance
may be achieved by simply immersing the tube in a coolant bath or inlet
temperature not to exceed 30°C. In demanding situations it is desirable
to circulate the liquid toward certain tube "hot spots" to
achieve a higher degree of cooling. To this end, Perkin Elmer offers
thyratrons with cooling pipes attached to the anode and/or grid area
to allow direct circulation of liquid for cooling.
Hollow Anode
Basic thyratrons are much like normal vacuum tubes in that they have
a heated cathode and a relatively cool flat, solid anode. The flat,
solid anode is not a good electron emitter and therefore the tube will
act like a diode when a reverse voltage (i.e., negative with respect
to the cathode) is applied to the anode (it will not conduct in the
reverse direction). This characteristic is desirable in most instances.
However, in some under-damped pulse circuits, the reverse (negative)
voltage on the flat, solid anode becomes high enough to cause conduction
for one or more reverse half-cycles of the discharge current waveform.
In those instances, anode damage is caused by cathode spot formation.
A cathode spot is a very localized molten spot that is high enough in
temperature to emit electrons. Unfortunately, the temperature of this
spot is usually high enough to liquefy the anode material and evaporate
it from the anode surface. Not only does this result in the loss of
anode material, but it usually results in deposition of some of that
material on the ceramic insulators of the tube. When the insulators
become locally conductive near the very high electric field space that
exists between the anode and the grid, arcing from anode to grid through
that locally conductive layer usually results and further leads to deteriorating
hold off performance of the tube.
To improve electron emission of the anode during reverse half cycles,
holes are made in the anode's flat face that lead to a cavity behind
that face. A significant number of electrons enter those holes on the
forward half cycle of the tube's conduction. These electrons are available
to contribute to reverse current flow during the reverse half-cycles
of the discharge waveform since many have not yet had time to become
trapped within the metal walls of the anode. Perkin Elmer hollow anode
thyratrons are usually rated to conduct in the reverse direction up
to 40% of the peak current that was conducted on the immediately previous
positive half cycle of the current discharge waveform. Higher reverse
current levels are achievable at the expense of some damage to the anode.
However, since the electron emission predominates from cathode spots
that are formed within the holes or cavity of the anode, the metal vapor
byproducts of those spots do not coat the ceramic insulator. Hollow
anode thyratrons do "wear" at a higher rate as the magnitude
of reverse current increases, but they do not generally exhibit the
rapidly deteriorating hold off capacity of solid anode tubes.
Grounded Grid
In applications with extremely high peak current and rate of rise of
current, a grounded grid thyratron is the best choice. The thermionic
cathode of the grounded grid thyratron only has to inject enough electrons
into the space between the grid baffle and the grid itself to start
the hollow cathode process that is described in most "pseudospark"
switch literature. Almost all of the electrons that are involved in
the grid-anode discharge are generated at cathode spots in the hollow
cathode area that is located in the grid slots and at the back surface
of the grid away from the grid face (the large flat area that directly
faces the anode). Proper design of the slot area, grid material, and
spacing of the grid from the anode will allow extremely high peak currents
at extremely short rise times while minimizing grid wear. Currently
grounded grid thyratrons all have had solid, flat anodes. Anode and
grid wear due to cathode spots has been the life-limiting factor in
the grounded grid tube. See the individual tube's data sheet for more
specific information.
Low Inductance
Perkin Elmer's low inductance tubes are designed to fit into very tight
discharge loop circuits. Of necessity, this generally means that they
have an overall seated height that is significantly shorter than the
length of a tube that has similar electrical parameters but one expected
to operate in a circuit with more inductance.
In essence, the low inductance tubes are just "normal" thyratrons
that have been severely repackaged to fit into the smallest height circuit
that is reasonably possible.
Because of the significant overall height reduction in the thyratrons,
the length of the high voltage stand-off ceramics may not be sufficient
for operation at normal ambient atmospheric conditions. Under normal
conditions, 10 kV per inch may be applied across the ceramics. Industry
standard de-rating data should be consulted for the voltage that may
be applied to the ceramic under other conditions. Most users of low
inductance tubes usually find it necessary to operate those tubes in
pressurized alternative gases or dielectric liquids.
Current Derating Curve
Perkin Elmer data sheets will generally list the maximum peak current
capability of the individual thyratron under a section titled "Absolute
Ratings." That peak current is usually specified at a reasonably
short duration current pulse width such as 250 nsec and it generally
remains the maximum peak current rating for pulses shorter than the
pulse width that is listed. However, as the pulse width is increased
in time, the peak current limit must be reduced. The formula for rating
the peak current capability of a thyratron at wider pulse widths is
as follows:
ibt
= ib3(3/tp)1/2
where ib3 is the peak current rating at the pulse width
listed for the absolute maximum limit; tp is the pulse width
of interest (in µS); ibt is the peak current rating
at the pulse width of interest.
Thyratron
Lines
1 inch thyratrons: The typical thyratron in this line is able to switch
100 A at up to 8 kV. "Low power" pulsers and Pockels cell
drivers are typical of the applications for these tubes.
1 1/2 inch thyratrons: The typical thyratron in this line is able to
switch 350 A at up to 16 kV. Many mechanical and a few electrical variations
of the basic tube design are readily available. Typical applications
are in Pockels cell drivers, radar transmitters, and portable gas lasers.
Product availability is subject to change without notice. For current
pricing and availability, please
contact your local Richardson sales office.
< < Back To the Hydrogen
Thyratrons Page |